What's in this template?
This Practice Equipment Register and Maintenance Log gives Australian general practices a complete equipment register aligned to RACGP Standards 5th edition practice equipment indicators (GP5.2 and GP5.3) and the relevant Australian Standards for fire equipment, emergency lighting, electrical testing and medical equipment. In the 6th edition, published on 26 August 2026, the same subject sits under criterion CG11.A (equipment for primary care, emergency care and resuscitation), which changes the maintenance interval and the equipment lists (see the FAQ below). The register covers both editions: a 12-month service and calibration rule for 6th edition practices, and a checklist of both editions' required equipment and doctor's bag lists. Accreditation assessments still use the 5th edition. The register is a fillable table with one row per item plus a covering policy that defines roles, intervals and audit cadence.
Important scope note: the vaccine fridge is not in this register, vaccine cold-chain management is covered by the dedicated Cold Chain Management policy and its own log. This register covers everything else.
The document covers 12 sections plus a sign-off block plus the Equipment Register table:
- Purpose: RACGP 5th edition GP5.2 and GP5.3, 6th edition CG11.A, and the WHS plant-and-equipment duty
- Scope: clinical, emergency, safety; excludes vaccine fridge and IT
- Roles and responsibilities: coordinator, manager, clinicians, contractors
- Register fields: item, location, asset ID, intervals, status
- Service and calibration intervals: the 5th edition rule (manufacturer's recommendations), the 6th edition rule (at least every 12 months), and an equipment-by-equipment table
- Electrical testing, AS/NZS 3760 and AS/NZS 3551: test and tag for non-medical equipment; medical electrical equipment tested and maintained under AS/NZS 3551 and the manufacturer's instructions
- Routine checks at every use: visual, tag in date, accessories, batteries
- The register: fillable table with starter rows for typical GP equipment, and a required equipment and doctor's bag checklist for both editions
- Faults, repair and decommissioning: out-of-service tagging, incident logging
- Audit: monthly coordinator, annual manager
- Related documents: links to Cold Chain Management for vaccine fridge
- Approval and review
Editable placeholder fields
- Practice name{{practice_name}}, Practice address{{practice_address}}
- Equipment coordinator{{equipment_coordinator}}, Practice manager{{practice_manager}}
- Electrical tester{{electrical_tester}}, your test-and-tag contractor
- Register rows for every piece of equipment
- Sign-off blocks for Practice Manager and Equipment Coordinator
Who needs a Practice Equipment Register?
Every Australian medical and allied health practice. RACGP accreditation looks for a documented register of clinical equipment with maintenance, calibration and electrical testing on schedule. The template is suitable for:
- General practices of any size, solo, group, corporate
- Specialist medical practices: adapt the equipment list to scope
- Day procedure clinics: extend with procedural equipment
- Allied health practices: adapt to discipline-specific equipment
- NDIS providers: assistive technology and shared equipment
Why the vaccine fridge is excluded
Vaccine cold-chain management is its own discipline with regulator-specific requirements:
- Twice-daily temperature monitoring
- Continuous data logging
- Defined excursion management protocol
- Vaccine wastage reporting
- AIR reporting obligations
These do not fit a generic equipment register. The Cold Chain Management policy handles all of this with its own dedicated log. This register cross-references but does not duplicate.
What's in the equipment intervals table
A complete service/calibration table covering common Australian general practice equipment:
| Category | Examples |
|---|---|
| Clinical assessment | Sphygmomanometer, ECG, spirometer, pulse oximeter, otoscope, dermatoscope, doppler |
| Treatment | Autoclave (AS 5369:2023), examination couch, examination lights, suction |
| Emergency | AED/defibrillator, oxygen cylinder, emergency drugs kit, BVM |
| Fire safety | Extinguishers (AS 1851 6-monthly tagging), fire blanket, smoke alarms |
| Lighting and exits | Emergency lighting and exit signs (AS 2293 6-monthly) |
| Electrical | Non-medical electrical equipment (AS/NZS 3760 test and tag); medical electrical equipment such as the ECG, AED and suction unit (AS/NZS 3551) |
The starter register table includes typical GP equipment so you can begin filling in immediately.
Australian Standards referenced
- AS 1851: Routine service of fire protection systems (extinguishers, fire blankets)
- AS 2293: Emergency escape lighting and exit signs (6-monthly testing)
- AS/NZS 3760: In-service safety inspection and testing of electrical equipment (test and tag), for non-medical electrical equipment
- AS/NZS 3551: Management programs for medical equipment (medical electrical equipment, which AS/NZS 3760 does not cover)
- AS 5369:2023: Reprocessing of reusable medical devices and other devices in health and non-health related facilities (autoclave). It replaced AS/NZS 4815:2006 in December 2023
- AS 2030: Gas cylinders periodic inspection
How to customise this template
- Download the Word document and replace every Placeholder{{placeholder}} with your details
- Walk the practice: every room, every drawer. List every piece of clinical and safety equipment in section 8 (Register). One row per item
- Add asset IDs: most practices use a simple format like
PE-001,PE-002 - Confirm intervals: defer to the manufacturer instructions; default to 12-monthly where the manufacturer is silent. On the 6th edition, service and calibrate every listed item at least every 12 months
- Schedule electrical testing: AS/NZS 3760 test and tag for non-medical electrical equipment, usually annually, by a licensed contractor; medical electrical equipment is tested and maintained under AS/NZS 3551 and the manufacturer's instructions
- Note any state variation: particularly for electrical testing in healthcare settings
- Set the AED monthly check: most practices nominate the practice nurse or a senior receptionist; visual indicator self-test, pad and battery expiry monitoring
- Tag faulty equipment immediately: "OUT OF SERVICE" and remove from clinical use
Related templates and tools
The Practice Equipment Register sits alongside several other practice safety documents:
- Electrical Safety and Test-and-Tag Policy: the AS/NZS 3760 risk classification and testing-interval rules behind this register's electrical section
- Cold Chain Management: vaccine fridge has its own dedicated log (not in this register)
- Infection Control Policy: autoclave validation and cycle records
- Emergency Response Plan and Emergency Evacuation Plan: fire equipment context
- Computer and Information Security Policy: separate IT asset register
- Incident and Hazard Report Form: equipment failure causing or potentially causing harm
- Quality Improvement Policy and Activity Log: register audit outputs
Frequently asked questions
Why isn't the vaccine fridge in this register?
The vaccine fridge has its own regulatory framework, twice-daily temperature monitoring, continuous data logging, defined excursion management, AIR reporting. It does not fit a general equipment register. The Cold Chain Management policy and its dedicated log cover it. This register notes the exclusion and points to that policy.
What service interval should we use?
Under the 5th edition, defer to the manufacturer's instructions for use and default to 12-monthly where the manufacturer is silent. Under the 6th edition (CG11.A), service and calibrate each listed item at least every 12 months, and sooner where the manufacturer says so. Some equipment has a tighter cadence (autoclave cycle records per cycle; spirometer daily calibration; AED monthly visual check). The table in section 5 lists the typical intervals.
What changes under the RACGP 6th edition?
The 5th edition asks for equipment to be maintained to the manufacturer's requirements. Criterion CG11.A of the 6th edition asks for each item to be maintained and calibrated annually, and its required practice equipment list and doctor's bag list differ from the 5th edition's. A practice working to the 6th edition should check the register against the 6th edition lists, which the RACGP publishes on its CG11 web page, and make sure each item on them has a service and calibration date no more than 12 months old, rather than using 12 months only where the manufacturer is silent. Section 8 of the register has a checklist of both editions' required equipment and doctor's bag lists, with the items that appear in only one edition marked, and section 5 sets out the 12-month rule.
Do we need test-and-tag and how often?
Yes for the practice's non-medical portable electrical equipment, typically annually under AS/NZS 3760; the Electrical Safety and Test-and-Tag Policy sets the intervals. Medical electrical equipment is different: Safe Work Australia's electrical Code of Practice notes that "AS/NZS 3760:2010 specifically excludes medical devices and electrical devices in patient care areas" and points to AS/NZS 3551 for medical equipment, so devices such as the ECG, AED and suction unit are inspected, tested and maintained as AS/NZS 3551 and the manufacturer's instructions require. Engage a licensed test-and-tag contractor for your jurisdiction; some practices choose 6-monthly for higher-use items.
What about the AED?
Monthly visual check by the practice (battery indicator, pad expiry); annual service per the manufacturer's instructions (the RACGP 6th edition guidance names defibrillators among the equipment to be serviced annually). Most AEDs have a self-test indicator that's checked daily or weekly. Replacing pads and battery on expiry is the most common ongoing cost.
How do we handle faulty equipment?
Tag "OUT OF SERVICE" immediately and remove from clinical use. The Equipment Coordinator investigates and decides on repair, replacement or decommission. Faults that caused or could have caused patient harm are also logged in the Incident and Hazard Report Form. The policy covers this in section 9.
Will accreditors accept this template?
Yes, when populated. The register maps to the RACGP 5th edition practice equipment indicators (GP5.2 and GP5.3), 6th edition criterion CG11.A and the relevant Australian Standards. Surveyors will spot-check items at random, pick up a sphygmomanometer, ask when it was last calibrated, look at the tag. A current register with up-to-date entries answers this well.
What's the relationship between this register and the autoclave cycle records?
This register notes the autoclave's annual validation and service. The infection control cycle records (the printout checked before each load is released, chemical indicators on every pack, and the load record, per AS 5369:2023 and the RACGP IPC guidelines) are kept under the Infection Control Policy. Both are needed; both are scrutinised at accreditation; both are referenced in the register entry.